FDA recall record

Abbott Molecular 5 mL Reaction Vessels recall — FDA Z-2169-2014

Abbott Molecular identified that some of the 5 ml Reaction Vessels (RVs) contained in lots 56944001, 57026001, 57059001, and 57673001 (original vendor lot…

Class IIclassification
Z-2169-2014official record ID
Aug 20, 2014report date

Recall record at a glance

Official recordZ-2169-2014
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAbbott Molecular
Report dateAug 20, 2014
Recall initiation dateJun 19, 2014
Quantity1,688,000 reaction vessels
LocationDes Plaines, IL, United States

Official product description

5 mL Reaction Vessels (a consumable of the Abbott m2000sp Automated Fluid Handling System). The Abbott m2000 RealTime System is intended for use in performing nucleic acid amplification testing by polymerase chain reaction in clinical laboratories. The system is comprised of the m2000rt and the m2000sp instruments. The Abbott m2000 RealTime System is intended for use in performing nucleic acid amplification testing by polymerase chain reaction in clinical laboratories. The system is comprised of the m2000rt and the m2000sp instruments.

Why the product was recalled

Abbott Molecular identified that some of the 5 ml Reaction Vessels (RVs) contained in lots 56944001, 57026001, 57059001, and 57673001 (original vendor lot 2198077) may have an insufficient rim. An RV with an insufficient rim may not be held securely in the 1 mL Subsystem Carrier. This may cause the RV to be seated incorrectly in the carrier or drop through it.

Distribution information

Worldwide Distribution - USA (nationwide) and Internationally to: AUSTRALIA AUSTRIA BELGIUM BENIN CAMEROON CHINA ETHIOPIA FINLAND FRANCE GERMANY INDIA INDONESIA IRELAND ISRAEL ITALY KENYA MALAYSIA MALAWI MALI NETHERLANDS NEW ZEALAND NORWAY PAKISTAN PHILIPPINES POLAND PORTUGAL RUSSIA SENEGAL SINGAPORE SOUTH AFRICA SOUTH KOREA SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UKRAINE UNITED KINGDOM VIETNAM

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.