Recall record at a glance
| Official record | Z-0699-2021 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | ICU Medical, Inc. |
| Report date | Dec 30, 2020 |
| Recall initiation date | Aug 31, 2020 |
| Quantity | 19850 |
| Location | San Clemente, CA, United States |
Official product description
42 IN (107 cm) Appx 5.0 ml, 20 Drop Admin Set, Spiros, REF: 20131-01; 31" (79 cm) Admin Set w/20 Drop Integrated Clave Drip Chamber, Spiros, Bag Hanger Drop-in Red Cap, 4 Units, REF: CH3330-C; 60 IN (152 cm) Appx 1.1 ml, Smallbore Set, MicroClave Clear, Spiros, REF: CH3353; 8" (20 cm) Appx 0.43 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, 2 Clamps, Spiros, REF: CH3367; 7" (18 cm) Appx 0.40 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, Spiros, 2 Clamps, REF: CH3379; 16" (41 cm) Appx 4.4 ml, Ext Set w/MicroClave, 0.2 Micron Filter, Clave, Spiros, Clamp, REF: CH3381; 31" (79 cm) Appx 3.3 ml, Admin Set w/20 Drop Integrated Clave Drip Chamber, Spiros w/Drop-in Red Cap, 2 Hanger, REF: CH3386; 40" (102 cm) Appx 7.3 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, 0.2 Micron Filter, Spiros, Hanger, REF: CH3401; 60" (152 cm) Appx 0.62 ml, Smallbore Ext Set w/MicroClave, Clamp, Spiros w/Red Cap, REF: CH3405; 75" (191 cm) Appx 9.1 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Clave, Spiros w/Red Cap, REF: CH3453; 73" (185 cm) Appx 2.1 ml, Smallbore Ext Set w/MicroClave Clear, Spiros, Clamp, REF: CH3485; 42" (107 cm) Appx 4.8 ml, 15 Drop Admin Set w/Spiros w/Red Cap, Bag Hanger, Drop-In Priming Cap, REF: CH3517; 8" (20 cm) Appx 0.49 ml, Smallbore Bifuse Ext Set w/MicroClave Clear, 2 Spiros, 3 Clamps, Priming Cap, REF: CH3523; 74" (188 cm) Smallbore Ext Set w/MicroClave, 0.2 Micron Low Protein Binding Filter, Spiros w/Red Cap, Clamp, REF: CH3533; 74 IN (188 cm) Appx 0.92 ml, Smallbore Set, MicroClave, 1.2 Micron Fil, Spiros, Red Cap, REF: CH3538; 18" (46 cm) Appx 5.1 ml, Ext Set w/0.2 Micron Filter, Clave, Spiros, Clamp, REF: CH3556; 60" (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave Clear, 0.2 Micron Low Protein Binding Filter, Spiros w/Red Cap, Clamp, REF: CH3603; 4-Way Clave Stopcock w/2 Spiros, REF: B4194; Spiros, Closed Male Connector w/Red Cap (10 Units), REF: Z7093; 47" (119 cm) 150 ml Burette Set w/MicroClave, 20 Drop In-Line Drip Chamber w/Ball Shut Off, Spiros, REF: Z7182
Why the product was recalled
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
Distribution information
Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.