FDA recall record

ICU Medical 3mL Syringe w/Spinning Spiros, Red Cap, REF: CH2003 recall — FDA Z-0704-2021

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion,…

Class IIclassification
Z-0704-2021official record ID
Dec 30, 2020report date

Recall record at a glance

Official recordZ-0704-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmICU Medical, Inc.
Report dateDec 30, 2020
Recall initiation dateAug 31, 2020
Quantity534764
LocationSan Clemente, CA, United States

Official product description

3mL Syringe w/Spinning Spiros, Red Cap, REF: CH2003; 5mL Syringe w/Spinning Spiros, Red Cap, REF: CH2005; 10mL Syringe w/Spinning Spiros, Red Cap, REF: CH2010; 20mL Syringe w/Spinning Spiros, Red Cap, REF: CH2020; 30mL Syringe w/Spinning Spiros, Red Cap, REF: CH2030; 60mL Syringe w/Spinning Spiros, Red Cap, REF: CH2060; 30" (76 cm) Appx 3.3 mL, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Red Cap, Bag Hanger, REF: CH3011; 31" (79 cm) Appx 3.3 ml, Admin Set w/20 Drop Integrated Clave Drip Chamber, Spiros w/Red Cap, Hanger, REF: CH3011-C; 40 IN (102 cm) App 4.8 ml, Admin Set, 20 Drop, 2 Spiros, Bag Hanger, REF: CH3020; 14 IN (36 cm) App 4.8 ml, Bifuse Add-On Set, Spiros w/Red Cap, Bag Spike, REF: CH3033; 5" (13 cm) Bag Spike Adapter w/Spiros w/Red Cap, Vented Cap, REF: CH3034; 30" (76 cm) Appx 5.9 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, 0.2 Micron Filter, Spiros w/Red Cap, Hanger, REF: CH3128; Oncology Kit w/60" (152 cm) Appx 2.3 ml, Smallbore Ext Set w/Spiros, Anti-Siphon Valve, ChemoClave, Clamp, Rotating Luer, Double ChemoClave Syringe Transfer Set, Spiros w/Red Cap, REF: CH3606; 5" (13 cm) Appx 0.50 ml, Smallbore Trifuse Ext Set w/3 ChemoClave (Purple, Green Rings), Spiros w/Red Cap, 3 Clamps, REF: CH3625; 86" (218 cm) Appx 10.4 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Red Cap, REF: CH3642; Oncology Kit w/40" (102 cm) Appx 4.7 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Red Cap, Bag Hanger, Double ChemoClave Syringe Transfer Set, REF: CH3677; 5.5" (14 cm) Appx 0.55 ml, Smallbore Trifuse Ext Set w/3 ChemoClave, Spiros w/Red Cap, 3 Clamps, REF: CH3700; 50 IN (127 cm) Appx 6.4 ml, Ext Set w/ChemoClave, Spiros w/Red Cap, Clamp, REF: CH3703; 16 IN (41 cm) Appx 1.6 ml, Ext Set, ChemoClave, Spiros w/Red Cap, REF: CH3715; Oncology Kit, Ext w/Spiros, Grad Conn, 2 Vial Spikes w/Clave, Spinning Spiros , REF: CH3929; Oncology Kit, w/Ext, MicroClave, Clave, Grad Conn, Transfer, Spiros, Vial Spike, Clave, REF: CH3932; Oncology Kit, 12 IN Ext, Spiros, Graduated Adapter, 2 ChemoClave w/Vial Spike, Spiros, REF: CH3943; Oncology Kit w/12" Ext Set w/Spiros w/Red Cap, Clamp, Graduated Adapter, ChemoClave Vented Vial Spike, 20mm, ChemoClave Vented Vial Spike, 13mm, REF: CH3944; Oncology Kit, 16 IN (41 cm) Ext Set, ChemoClave, Graduated Conn, 2 ChemoClave w/Vial Spike, Spiros, REF: CH3946; Oncology Kit, 21" (53 cm) Appx 3.2 ml, Ext Set w/Spiros w/Red Cap, Graduated Adapter, Locking Universal Vented Vial Spike w/ChemoClave, Closed Vial Spike, 13mm w/ChemoClave, Spinning Spiros w/Red Cap, REF: CH3949; 46" (117 cm) 50 ml, Diluent Set for Channel 2 w/20 Drop Drip Chamber, Spiros, REF: CH4002

Why the product was recalled

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

Distribution information

Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceSep 23, 2026

ICU Medical Plum administration sets recall — FDA Z-3156-2026

Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the…

Class IIFDA: OngoingICU Medical, Inc.