FDA recall record

PIE Medical Imaging B.V 3mensio Workstation (Vascular Fenestrated) software recall — FDA Z-2767-2026

User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%)…

Class IIclassification
Z-2767-2026official record ID
Jul 29, 2026report date

Recall record at a glance

Official recordZ-2767-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmPIE Medical Imaging B.V.
Report dateJul 29, 2026
Recall initiation dateMay 7, 2026
Quantity165
LocationMaastricht, N/A, Netherlands

Official product description

3mensio Workstation (Vascular Fenestrated) software

Why the product was recalled

User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the following occur: affected software is used, and the user has access only to 3D Echo OR Mitral 3D Echo, and the dataset being analyzed is a Philips 3D echo dataset containing Color Doppler information

Distribution information

Worldwide - US Nationwide distribution in the states of CA, NY, MN, MI, TX and the countries of United Kingdom, Australia, France, Italy, Hong Kong SAR, Thailand, Taiwan, Netherlands, Ireland, Spain, Greece, Japan, Belgium.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: N/A.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.