Recall record at a glance
| Official record | Z-2517-2023 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Completed |
| Recalling firm | Philips Ultrasound, Inc. |
| Report date | Sep 13, 2023 |
| Recall initiation date | Jun 30, 2023 |
| Quantity | 449 |
| Location | Bothell, WA, United States |
Official product description
3D9-3v Transducer, accessory to Philips Ultrasound Systems: EPIQ Elite; Affiniti 30, 50, and 70; ClearVue 850; HD15; iU22; and Compact 5000
Why the product was recalled
Transducer, an ultrasound system accessory, consists of two parts that are bonded together that may come apart due to chassis bonding issue, which may pose risk to users and patients of tissue damage, and electric shock if the built-in safety measures to prevent electric were to also fail.
Distribution information
Worldwide - US Nationwide distribution including in the states of FL, MN, AZ, NC, CA, MO, OH, WV, AL, NH, IN, IL, VT, PA, NY, NE, TX, WI, MI, KY, NV, MA, SD, LA, CO, TN, AK, GA, WA, UT, NM, ME, AR, SC and the countries of Italy, China, Netherlands, New Zealand, Hungary, Australia, France, Thailand, C¿te D'Ivoire, Germany, Canada, Mexico, Spain, Saudi Arabia, Uzbekistan, Greece, Yemen.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.