Recall record at a glance
| Official record | Z-2837-2018 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Siemens Medical Solutions USA, Inc. |
| Report date | Aug 29, 2018 |
| Recall initiation date | Jun 29, 2018 |
| Quantity | 699 |
| Location | Malvern, PA, United States |
Official product description
3D TOP Ceiling Stand, model no. 3070039 3074486 3070013 3070021 Product Usage: The ceiling stand is used to hold the X-ray tube and / or the detector and is moveable on a ceiling rail system in three axis. The tube and / or detector can be rotated above a horizontal and a vertical axle. The ceiling stand can be used within a stationary diagnostic X-ray system as well.
Why the product was recalled
In rare cases of insufficient maintenance or high clinical workload, the first rope of the ceiling stand, which is designed to take the load, could break without triggering the safety lock and this could lead to overloading the safety rope which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down during patient positioning. Patients and users may be seriously injured.
Distribution information
US Nationwide Distribution to accounts in IL, FL, and NY.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.