FDA recall record

Trilliant Surgical 2.4, 3.0, and 4.0 mm Gridlock Screw Driver Bit, Model Number 310-30-003 recall — FDA Z-2738-2015

Gridlock screw driver bits for the Gridlock Plating System were identified to be out of specification and have the potential for shallow engagement…

Class IIclassification
Z-2738-2015official record ID
Sep 16, 2015report date

Recall record at a glance

Official recordZ-2738-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmTrilliant Surgical Ltd.
Report dateSep 16, 2015
Recall initiation dateJul 31, 2015
Quantity108 units
LocationHouston, TX, United States

Official product description

2.4, 3.0, and 4.0 mm Gridlock Screw Driver Bit, Model Number 310-30-003, Lot Number TSL002605, NON-STERILE. For use with the Gridlock Plating System for trauma and reconstructive procedures of small bones.

Why the product was recalled

Gridlock screw driver bits for the Gridlock Plating System were identified to be out of specification and have the potential for shallow engagement between the driver and the screw, resulting in an increased likelihood of stripping screw heads intraoperatively. Specifically, the minor diameter of the drive feature is above the specification, potentially reducing the engagement between the driver and the screw head.

Distribution information

NY, OH, CA, SC, NJ, AZ, VA, IN, FL, MN, OK, AR, MO, WI, GA, SD, IA, IL, TX

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.