FDA recall record

Merit Medical Systems 10Fore Hemostasis Valve Catalog Number / UDI-DI code: MAPTEN4 / recall — FDA Z-2553-2025

Due to a manufacturing issue involving the silicone quad ring where the ring is deformed or otherwise damaged and may cause a foreign body in the fluid…

Class IIclassification
Z-2553-2025official record ID
Sep 17, 2025report date

Recall record at a glance

Official recordZ-2553-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMerit Medical Systems, Inc.
Report dateSep 17, 2025
Recall initiation dateJul 25, 2025
Quantity447
LocationSouth Jordan, UT, United States

Official product description

10Fore Hemostasis Valve Catalog Number / UDI-DI code: MAPTEN4 / 00884450821305 MAPTEN42 / 00884450832509 MAPTEN452 / 00884450821329 The 10Fore Hemostasis Valve is intended to maintain hemostasis during the introduction/withdrawal and use of diagnostic and interventional devices.

Why the product was recalled

Due to a manufacturing issue involving the silicone quad ring where the ring is deformed or otherwise damaged and may cause a foreign body in the fluid pathway.

Distribution information

Worldwide - U.S. Nationwide distribution in the states of CA, FL, LA, MT, NY, SD, and WI. The countries of Canada and Korea

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.