FDA recall record

Horiba Instruments, Inc dba Horiba Medical 10 ML REAGENT CUP, B1034626, SAP 1221034626 for recall — FDA Z-0493-2018

The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient…

Class IIclassification
Z-0493-2018official record ID
Feb 7, 2018report date

Recall record at a glance

Official recordZ-0493-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmHoriba Instruments, Inc dba Horiba Medical
Report dateFeb 7, 2018
Recall initiation dateNov 21, 2017
Quantity10 bags, 100 units each
LocationIrvine, CA, United States

Official product description

10 ML REAGENT CUP, B1034626, SAP 1221034626 for use with ABX PENTRA 400 / PENTRA C400

Why the product was recalled

The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.

Distribution information

Nationwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.