FDA recall record

EBI Patient Care 10-1335M - SpF-XL IIB 2/DM Implantable Stimulator recall — FDA Z-2118-2017

Zimmer Biomet is conducting a medical device field action for the SpF PLUS-Mini and SpF XL IIB Implantable Spinal Fusion Stimulator products due to higher…

Class Iclassification
Z-2118-2017official record ID
May 31, 2017report date

Recall record at a glance

Official recordZ-2118-2017
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmEBI Patient Care, Inc.
Report dateMay 31, 2017
Recall initiation dateApr 20, 2017
Quantity33 units total
LocationGuaynabo, PR, United States

Official product description

10-1335M - SpF-XL IIB 2/DM Implantable Stimulator. Indicated as a spinal fusion adjunct to increase the probability of fusion success in 1 or 2 levels.

Why the product was recalled

Zimmer Biomet is conducting a medical device field action for the SpF PLUS-Mini and SpF XL IIB Implantable Spinal Fusion Stimulator products due to higher than allowed cytotoxicity levels found during routine process monitoring with the product.

Distribution information

Nationwide distribution only.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.