FDA recall record

Daavlin Distributing Company 1 Series CX phototherapy units equipped with Daavlin's recall — FDA Z-1073-2022

Software issue, resuming an interrupted treatment will result in swap of utilized calibration treatment distance 0" inches to "9. Resulting in over or…

Class IIclassification
Z-1073-2022official record ID
May 18, 2022report date

Recall record at a glance

Official recordZ-1073-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmDaavlin Distributing Company
Report dateMay 18, 2022
Recall initiation dateApr 26, 2022
Quantity216
LocationBryan, OH, United States

Official product description

1 Series CX phototherapy units equipped with Daavlin's ClearLink Control System software versions v3.03h and 3.04v. Models: [1 Ser CX 305-4 120-240V 50/60Hz, item # 06OS0004CX6], [1 Ser CX 311-4 120-240V 50/60Hz CE, item # 807OS0004CX5], [1 Ser CX 311-4 120-240V 50/60Hz, item # 807OS0004CX6], [1 Ser CX 350-4 120-240V 50/60Hz, item # 807OS0400CX6], [1 Ser CX UVA1-4 120-240V 50/60Hz, item # 807OS0400CX6A1]

Why the product was recalled

Software issue, resuming an interrupted treatment will result in swap of utilized calibration treatment distance 0" inches to "9. Resulting in over or under dosing.

Distribution information

Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, VA, WA, WI, WV and the countries of Australia, Canada.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Telephone.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.