FDA recall record

Acumed .045 X 6 ST Guide Wire, Part number 45-0029-S, Sterile 10 pack of WS-1106ST and has a recall — FDA Z-1483-2017

Guide wires sized as .045 X 6 were mis-packaged as .054 X 7 guide wires and those sized as .054 X 7 were mis- packaged as .045 X 6 guide wires. There is a…

Class IIclassification
Z-1483-2017official record ID
Mar 22, 2017report date

Recall record at a glance

Official recordZ-1483-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAcumed LLC
Report dateMar 22, 2017
Recall initiation dateFeb 17, 2017
Quantity24 units of .045 X 6 ST Guide Wire and 25 units of .054 X 7 ST Guide Wire
LocationHillsboro, OR, United States

Official product description

.045 X 6 ST Guide Wire, Part number 45-0029-S, Sterile 10 pack of WS-1106ST and has a 6 inch length and 0.045 inch diameter. .054 X 7 ST Guide Wire, Part number 45-0031-S, Sterile 10 Pack of WS-1407ST and has a 7 inch length and 0.054 inch diameter. These guide wires are an orthopedic manual surgical instrument.

Why the product was recalled

Guide wires sized as .045 X 6 were mis-packaged as .054 X 7 guide wires and those sized as .054 X 7 were mis- packaged as .045 X 6 guide wires. There is a possibility that using the incorrect guide wire size can cause health hazards, such as mal or nonunion, need for revision surgery, and histological response (if wire breaks).

Distribution information

Worldwide Distribution in Great Britain, Ireland, and Japan only.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.