Recall record at a glance
| Official record | D-0080-2025 |
|---|---|
| Category | Drug Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | American Regent, Inc. |
| Report date | Dec 11, 2024 |
| Recall initiation date | Nov 11, 2024 |
| Location | Shirley, NY, United States |
Official product description
Venofer (iron sucrose) Injection, USP 100 mg Elemental Iron per 5 mL (20 mg/mL), 5 mL Single-Dose Vials, Rx Only, For Intravenous Use Only, Distributed by: Fresenius Medical Care NA, Waltham, MA 02451, NDC: 49230-534-01 (vial), NDC: 49230-534-25 (25 x 5 mL/vial cartons).
Why the product was recalled
Presence of Particulate Matter: Potential for glass delamination from the vials.
Distribution information
Nationwide in the USA
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.