Recall record at a glance
| Official record | D-0525-2026 |
|---|---|
| Category | Drug Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Oasis Medical, Inc. |
| Report date | May 20, 2026 |
| Recall initiation date | Apr 27, 2026 |
| Location | Glendora, CA, United States |
Official product description
Oasis Tears PF, Preservative-Free Lubricant Eye Drops, 10mL/0.34 Fl OZ Bottle, Sterile, Manufactured for: OASIS Medical, Inc., 514 S. Vermont Ave., Glendora, CA 91741, USA, Made in France, NDC 42126-6400-1.
Why the product was recalled
Lack of Assurance of Sterility: The recall is being initiated out of an abundance of caution following FDA observations noted during a recent inspection of Excelvision.
Distribution information
Nationwide within the United States
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.