FDA recall record

CIPLA Azacitidine for Injection 100 mg/vial For Subcutaneous and Intravenous Use Only recall — FDA D-0165-2021

cGMP deviations: Vials may not be sealed correctly affecting sterility.

Class IIclassification
D-0165-2021official record ID
Dec 23, 2020report date

Recall record at a glance

Official recordD-0165-2021
CategoryDrug Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCIPLA
Report dateDec 23, 2020
Recall initiation dateDec 14, 2020
Quantity2151 vials
LocationWarren, NJ, United States

Official product description

Azacitidine for Injection 100 mg/vial For Subcutaneous and Intravenous Use Only Single-Dose Vial Manufactured by Shilpa Medicare Limited Jadcherla -509301 INDIA Distributed by: Cipla USA, Inc. 10 Independence Blvd, Suite 300 Warren, NJ 07059 NDC 69097-805-40

Why the product was recalled

cGMP deviations: Vials may not be sealed correctly affecting sterility.

Distribution information

USA Nationwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.