Monitor U.S. medical device recall enforcement reports from the FDA. Review product descriptions, recall reasons, classifications, dates, recalling firms and distribution information where the FDA provides them.
25,676medical device recall records
2,265FDA Class I records
8firms with 5+ records
Zimmer Biomet, Inc.759 records
Database records
Records — page 616
25,676 detailed records are currently available in this category.
There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop…
The firm has received reports of tip separation during usage of the catheters, which may expose patients to unintended dislodgement of the tip from the body of the catheter.
Class IFDA: TerminatedApplied Medical Resources Corp
There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop…
There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop…
The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to…
Shipments were transported outside of the required refrigerated temperature conditions. Potential hazards if used include nerve damage resulting in paralysis and increased potential for…
Packages of abutments may contain an incorrect screw. While the screw will tighten, it cannot be guaranteed that the screw will not loosen. In case the screw loosens the abutment should be…
To reinforce existing instructions within the Directions for Use (DFU) and provide further guidance which will be added to the DFU that may reduce the likelihood of metal cap and fiber tip…
Class IIFDA: TerminatedBoston Scientific Corporation
When a user acquires XperCT scan on an Azurion 2.0 system, enters the XperGuide guidance step and moves the L-arm away from the initial scan position before starting the live guidance, a…
Potential issue with the manufacture and inspection of U-channel sheaths in Implantation Instruments Sets. Affected U-channel sheaths may be more likely to cause damage to ProACT implant…
Registering dose and structures in the treatment delivery workflow could result in an alignment discrepancy between the imported previously delivered dose and the displayed patient anatomy…
During the sterilization of radial compression devices, some of the inflatable balloons may break or burst, which may result in air leakage and the inability to maintain…
Class IIFDA: TerminatedMerit Medical Systems, Inc.
The vertical dimensions of screw-hole position in some standalone cages might be out of range in engineering drawings. A possibility exists of screw pullout, lossening, instability, or less…
A faulty locking piece of the Instrument Carrier that does not fulfil the requirement of locking force to the Arc has been found during an internal check of a Leksell Vantage Stereotactic…
Class IIFDA: TerminatedElekta Inc.
How to use this medical device recall index
Medical device recall records can change as the FDA receives additional information. Check the official record identifier and current FDA information before making health or product-use decisions.
Recall Lookup USA keeps the last complete source record available when an official data service is temporarily unavailable. When a source provides only a few fields, we show only those known facts rather than padding the page with generic text.
What the database can and cannot tell you
This database is designed for discovery and reference. It can help you locate a relevant recall record, identify the recalling firm or NHTSA campaign, compare dates and classifications, and follow the official source. It is not a substitute for a VIN-specific NHTSA check, a manufacturer notice, medical advice or instructions issued by the recalling company.