Sonendo GENTLEWAVE REF: FG-002-0001 recall — FDA Z-2162-2020
The console would continue to run for extended period when the foot pedal was released.
Monitor U.S. medical device recall enforcement reports from the FDA. Review product descriptions, recall reasons, classifications, dates, recalling firms and distribution information where the FDA provides them.
25,676 detailed records are currently available in this category.
The console would continue to run for extended period when the foot pedal was released.
The firm is initiating a voluntary removal of certain spinal cord stimulation leads, lead extensions, adapters, and deep brain stimulation lead extensions due to the "use by date" displayed…
The unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an…
The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms…
Potential presence of elevated endotoxin levels that exceed the specification limit
Customers using the firm's continuous glucose monitoring system application on an iPhone, iPad or iPod Touch with ios software version 12, 12.1, or 12,2 are likely to experience a shortened…
Teleflex received the attached Medical Device Recall Notice from Pajunk Medical. Teleflex Medical is a distributor of Pajunk products. According to the notice, Pajunk internally identified…
Assigned Rheumatoid Factor (RF) calibrator value for calibrator component included in kit used to aid in the diagnosis of rheumatoid arthritis over recovers by 7% in comparison to the…
The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms…
The identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation…
The firm is initiating a voluntary removal of certain spinal cord stimulation leads, lead extensions, adapters, and deep brain stimulation lead extensions due to the "use by date" displayed…
Teleflex received the attached Medical Device Recall Notice from Pajunk Medical. Teleflex Medical is a distributor of Pajunk products. According to the notice, Pajunk internally identified…
If the C-arm leaves its intended travel path due to a fault within the drivetrain, movements may be impacted or impossible and the area of interest cannot be reached. In the event the…
Abbott identified that samples tested for ARCHITECT STAT High Sensitive Troponin I and STAT High Sensitivity Troponin I assays (LN 3P25 and LN 2R98) may show interaction when processed…
The rotary therapy selector switch may fail, resulting in unexpected device behavior including: 1) The device may not turn on; 2) The device may not perform the selected function;3) The…
There is a remote potential for the sterile packaging of lots of the transseptal guiding sheath to be compromised during shipping and handling. In the case that a packaging defect is not…
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
Three products requiring refrigerated storage conditions were incorrectly stored outside of approved temperature ranges. Using these temperature abused products may result in misdiagnosis…
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
Failure of cardiovascular clips to provide sufficient occlusion of the vessel.
Potential product mix-up. The recalling firm has determined that one unit of OsteoVation EX, 3cc that may not have been sterilized was inadvertently packaged as CranioSculpt C, 10cc.
The lot number on the pediatric orthopedic implant driver instrument handle does not match the lot number on the label.
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
Loose silicone particulate was found to be present on the shaft of the silicone catheters.
Medical device recall records can change as the FDA receives additional information. Check the official record identifier and current FDA information before making health or product-use decisions.
Recall Lookup USA keeps the last complete source record available when an official data service is temporarily unavailable. When a source provides only a few fields, we show only those known facts rather than padding the page with generic text.
This database is designed for discovery and reference. It can help you locate a relevant recall record, identify the recalling firm or NHTSA campaign, compare dates and classifications, and follow the official source. It is not a substitute for a VIN-specific NHTSA check, a manufacturer notice, medical advice or instructions issued by the recalling company.