Monitor U.S. medical device recall enforcement reports from the FDA. Review product descriptions, recall reasons, classifications, dates, recalling firms and distribution information where the FDA provides them.
24,239medical device recall records
2,205FDA Class I records
8firms with 5+ records
Zimmer Biomet, Inc.751 records
Database records
Records — page 439
24,239 detailed records are currently available in this category.
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display…
Class IIFDA: OngoingSiemens Medical Solutions USA, Inc.
Non-EUA authorized, CE marked, SARS-CoV-2 antigen rapid test kits are being recalled from the U.S. market because use may lead to inaccurate test results, including false negative or false…
Class IIFDA: OngoingAcon Biotech (Hangzhou) Co., Ltd.
Cannulated Screws 4.0mm x 10mm and 4.0mm x 12mm may not have sufficient axial pullout strength for the indicated use. Remote possibility of resulting in a mal or nonunion of the bone…
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display…
Class IIFDA: OngoingSiemens Medical Solutions USA, Inc.
Due to a programmer software anomaly under very specific circumstance when executing a pacing capture Decrement Test in-clinic on implantable cardioverter and cardiac resynchronization…
Class IIFDA: OngoingSt. Jude Medical, Cardiac Rhythm Management Division
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display…
Class IIFDA: OngoingSiemens Medical Solutions USA, Inc.
Due to a programmer software anomaly under very specific circumstance when executing a pacing capture Decrement Test in-clinic on implantable cardioverter and cardiac resynchronization…
Class IIFDA: OngoingSt. Jude Medical, Cardiac Rhythm Management Division
Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth…
Class IIFDA: OngoingSt. Jude Medical, Cardiac Rhythm Management Division
As a result of Medtronic's integrated power console (authentication number: 222ACBZX00018000) being used with software version: v2.7.3.0, it was discovered that there was a defect that…
Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or…
Reports have been received that after normal or routine lateral forces are applied to the head fixation device, the split block clamp may deflect (rotationally around the axis) and not…
Labeling change to due to the risk of severe anaphylactoid reactions including shock in patients treated with the Liposorber LA-15 LDL Adsorption Column under concomitant ACE inhibitor…
Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or…
Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery
Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery
Ventilator mounting brackets for the Vivo 50/60/65 and Vivo 45LS ventilator models can become detached from the Breas Ventilator Trolley, this may cause the ventilator to fall from the…
Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the…
Class IIFDA: TerminatedGraphic Controls Acquisition Corporation
How to use this medical device recall index
Medical device recall records can change as the FDA receives additional information. Check the official record identifier and current FDA information before making health or product-use decisions.
Recall Lookup USA keeps the last complete source record available when an official data service is temporarily unavailable. When a source provides only a few fields, we show only those known facts rather than padding the page with generic text.
What the database can and cannot tell you
This database is designed for discovery and reference. It can help you locate a relevant recall record, identify the recalling firm or NHTSA campaign, compare dates and classifications, and follow the official source. It is not a substitute for a VIN-specific NHTSA check, a manufacturer notice, medical advice or instructions issued by the recalling company.