DePuy Orthopaedics GLOBAL UNITE STD STEM SZ 14 Product Code: 110014100 recall — FDA Z-0489-2022
Stems may have an undersized spigot bore diameter therefore separating the proximal body from the stem may not be possible without surgical delay and/or bone damage