K2M recall — FDA Z-2021-2020
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
Browse the newest complete recall records currently stored from FDA enforcement reports and NHTSA vehicle campaigns. “Latest” refers to the newest source-dated record in this database, not the timestamp of our last API check.
The newest recall records with useful source details are listed here.
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
There is a remote potential for the sterile packaging of the transseptal guiding sheath to be compromised during shipping and handling. In the case that a packaging defect is not identified…
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
A software issue that may affect the analyzer's result accuracy.
May exhibit increased levels of artifacts which could be difficult to distinguish from gram positive cocci, increased levels of artifacts could result in a false-positive Gram stain
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
Loose silicone particulate was found to be present on the shaft of the silicone catheters.
Loose silicone particulate was found to be present on the shaft of the silicone catheters.
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
The lot number on the pediatric orthopedic implant driver instrument handle does not match the lot number on the label.
Potential product mix-up. The recalling firm has determined that one unit of OsteoVation EX, 3cc that may not have been sterilized was inadvertently packaged as CranioSculpt C, 10cc.
Failure of cardiovascular clips to provide sufficient occlusion of the vessel.
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
Three products requiring refrigerated storage conditions were incorrectly stored outside of approved temperature ranges. Using these temperature abused products may result in misdiagnosis…
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
There is a remote potential for the sterile packaging of lots of the transseptal guiding sheath to be compromised during shipping and handling. In the case that a packaging defect is not…
The rotary therapy selector switch may fail, resulting in unexpected device behavior including: 1) The device may not turn on; 2) The device may not perform the selected function;3) The…
Potential to be contaminated with mold