TEM Systems ex-TEM Assay for ROTEM delta Thromboelastometry System recall — FDA Z-0302-2017
Customers complained of low/empty reagent vials.
Overview of recall records associated with TEM Systems Inc., including dates, categories and classifications where available.
Classification mix: Class II: 2.
Customers complained of low/empty reagent vials.
Customer complaints about failed target ranges for alpha angle. Firm has not completed the CAPA identifying the root cause. Projected completion of the CAPA by the firm is July 2016.
This page summarizes recall records in which the official source identifies TEM Systems Inc. as the recalling firm or manufacturer. The number of records should not be treated as a safety rating or a measure of company size, because recall counts depend on product mix, reporting periods and how government records are organized.