Tec Com Gmbh TECOTHERM NEO, 100-130/200-240 VAC, 50-60 Hz/ max. 350VA, Class I, Schutzklasse recall — FDA Z-0362-2020
The action is being initiated due to potential installation of faulty peltier-elements within the thermo-regulation system. Failure of the peltier-element would report a system failure and…