Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of…
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of…
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
out-of-specification sodium azide concentration in Syva Emit Tox Serum Tricyclic Antidepressant Calibrator/Controls Lot U2. While the assay results were not impacted, the overall health…
Class IIIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
Processing of urine albumin assay on the CI analyzers may lead to the following issues: Falsely depressed Atellica CH Urea Nitrogen (UN_c) results; Falsely depressed Atellica CH Enzymatic…
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
Out-of-specification sodium azide concentration in Syva Emit Disopyramide Enzyme Immunoassay Calibrator Lot U1. While the assay results were not impacted, the overall health risk to lab…
Class IIIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
The potential for intermittent well-to-well bias on the A1c_E assay on the Atellica CH and Atellica CI analyzer which could cause falsely depressed or falsely elevated HbA1c patient results.
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality…
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient…
Class IIFDA: OngoingSiemens Healthcare Diagnostics Inc
A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by…
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and…
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in…
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in…
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in…
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
The potential for falsely depressed Intact PTH patient results at the low end of the assay range, less than or equal to 50 pg/mL (less than or equal to 5.3 pmol/L), when using specific lots…
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias…
Class IIFDA: OngoingSiemens Healthcare Diagnostics Inc.
Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium,…
Class IIFDA: OngoingSiemens Healthcare Diagnostics Inc.
Contamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride.
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability…
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
Understanding this recall history
This page summarizes recall records in which the official source identifies Siemens Healthcare Diagnostics, Inc. as the recalling firm or manufacturer. The number of records should not be treated as a safety rating or a measure of company size, because recall counts depend on product mix, reporting periods and how government records are organized.
Compare the dates, categories and classifications together rather than relying on the total alone. Open an individual record for the recalled product, reason, distribution details and current instructions available from the source.