Seiler Instr. & Mfg. Co Alpha Air 6 ENT Microscope Microscope examination of the human ear recall — FDA Z-1294-2018
The microscope may tip over if attachments are fully extended in one line and there is an attempt to move the device not by the handles mounted on the post.