REPLIGEN MAVERICK Measurement Module Spectrometer PN 820-01367 recall — FDA Z-1024-2026
Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the…