NOBEL BIOCARE SERVICES AG NobelReplace Conical Connection (CC) PMC RP 4.3x10mm, REF: 37291 recall — FDA Z-2828-2020
Dental implant cap label may incorrectly indicate implant measurements 5.0x8mm. Actual implant measurements are 4.3x10mm. If not noticed prior to implant placement, this could lead to: 1) A…