Lab Vision CEA / CD66e Ab-2, Product code RB-368-R7, RB-368-R1 recall — FDA Z-0815-2016
Certain lots of CEA/CD66e Ab2 were produced with a dilution factor of 1:100 instead of the recommended 1:250.
Overview of recall records associated with Lab Vision Corporation, including dates, categories and classifications where available.
Classification mix: Class III: 2.
Certain lots of CEA/CD66e Ab2 were produced with a dilution factor of 1:100 instead of the recommended 1:250.
It was discovered that the labeling of MS-230 antibody incorrectly describes the concentration - label reads 0.2mg/ml but it is actually 0.4 mg/ml.
This page summarizes recall records in which the official source identifies Lab Vision Corporation as the recalling firm or manufacturer. The number of records should not be treated as a safety rating or a measure of company size, because recall counts depend on product mix, reporting periods and how government records are organized.
Compare the dates, categories and classifications together rather than relying on the total alone. Open an individual record for the recalled product, reason, distribution details and current instructions available from the source.