Fresenius Medical Care Renal Therapies Group, LLC Recalls
Overview of recall records associated with Fresenius Medical Care Renal Therapies Group, LLC, including dates, categories and classifications where available.
Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
Class IIFDA: OngoingFresenius Medical Care Renal Therapies Group, LLC
Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
Class IIFDA: OngoingFresenius Medical Care Renal Therapies Group, LLC
Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
Class IIFDA: OngoingFresenius Medical Care Renal Therapies Group, LLC
Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
Class IIFDA: OngoingFresenius Medical Care Renal Therapies Group, LLC
Additional Instructions: step-by step handling and inserting the cassette into the Liberty Cycler set. A damaged cassette can cause peritoneal dialysis (PD) solution to leak into the cycler…
Class IIFDA: TerminatedFresenius Medical Care Renal Therapies Group, LLC
In the hemodialysis machine during the cleaning/disinfection program, when there is a power failure or interruption, the dialysis program button should be greyed out when the machine is…
Class IIFDA: TerminatedFresenius Medical Care Renal Therapies Group, LLC
In the hemodialysis machine during the cleaning/disinfection program, when there is a power failure or interruption, the dialysis program button should be greyed out when the machine is…
Class IIFDA: TerminatedFresenius Medical Care Renal Therapies Group, LLC
The recalling firm identified a software issue related to the Patient Line Check (PLC) which may result in an increased risk of Overfill (also known as Increased Intraperitoneal Volume,…
Class IIFDA: TerminatedFresenius Medical Care Renal Therapies Group, LLC
When the recirculation ultrafiltration (UF) Goal is set to a value greater than 200ml in service mode and the user starts a treatment using the SLED (Sustained Low Efficiency Dialysis)…
Class IIFDA: TerminatedFresenius Medical Care Renal Therapies Group, LLC
The label specifies the potassium concentration is 2.0 mEq/L. Some of the product in the affected lot was found to have a potassium concentration of 2.25 mEq/L.
Class IIFDA: TerminatedFresenius Medical Care Renal Therapies Group, LLC
While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed…
Class IIFDA: TerminatedFresenius Medical Care Renal Therapies Group, LLC
When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed…
Class IIFDA: TerminatedFresenius Medical Care Renal Therapies Group, LLC
When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed…
Class IIFDA: TerminatedFresenius Medical Care Renal Therapies Group, LLC
When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed…
Class IIFDA: TerminatedFresenius Medical Care Renal Therapies Group, LLC
When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed…
Class IIFDA: TerminatedFresenius Medical Care Renal Therapies Group, LLC
Use of the Blood Pressure Graph located on the Crit-Line screen of the 2008T Hemodialysis Machine may result in a frozen screen, possibly causing the blood pump to stop.
Class IIFDA: TerminatedFresenius Medical Care Renal Therapies Group, LLC
Added Warning Statement: Fresenius 2008 Series Hemodialysis Machines False Blood Leak Alarm When Dialyzing Patients Treated with Hydroxocobalamin (or any form of Vitamin B-12)
Class IIFDA: TerminatedFresenius Medical Care Renal Therapies Group, LLC
On the label calcium (Ca) listed in the right upper corner of the label (red background) incorrectly indicates the product contains 2.5 mEq/L Ca. The concentration indicated in the list of…
Class IIFDA: TerminatedFresenius Medical Care Renal Therapies Group, LLC
The door latch may not fully close and the door may open unexpectedly. The door may make physical contact with anything near the door of the machine, including the patient. May cause…
Class IIFDA: TerminatedFresenius Medical Care Renal Therapies Group, LLC
Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to…
Class IIFDA: TerminatedFresenius Medical Care Renal Therapies Group, LLC
Understanding this recall history
This page summarizes recall records in which the official source identifies Fresenius Medical Care Renal Therapies Group, LLC as the recalling firm or manufacturer. The number of records should not be treated as a safety rating or a measure of company size, because recall counts depend on product mix, reporting periods and how government records are organized.
Compare the dates, categories and classifications together rather than relying on the total alone. Open an individual record for the recalled product, reason, distribution details and current instructions available from the source.