Recall company profile

Fresenius Medical Care Renal Therapies Group, LLC Recalls

Overview of recall records associated with Fresenius Medical Care Renal Therapies Group, LLC, including dates, categories and classifications where available.

26recall records
Sep 9, 2026latest record date
Nov 25, 2015earliest record date

Recall history in this database

Classification mix: Class II: 25, Class III: 1.

Medical DeviceJan 25, 2017

Fresenius 2008T Series Hemodialysis System recall — FDA Z-1026-2017

When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed…

Class IIFDA: TerminatedFresenius Medical Care Renal Therapies Group, LLC
Medical DeviceJan 25, 2017

Fresenius 2008K Series : Hemodialysis System recall — FDA Z-1027-2017

When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed…

Class IIFDA: TerminatedFresenius Medical Care Renal Therapies Group, LLC
Medical DeviceMar 2, 2016

Fresenius Naturalyte Liquid Acid Concentrate recall — FDA Z-0887-2016

On the label calcium (Ca) listed in the right upper corner of the label (red background) incorrectly indicates the product contains 2.5 mEq/L Ca. The concentration indicated in the list of…

Class IIFDA: TerminatedFresenius Medical Care Renal Therapies Group, LLC

Understanding this recall history

This page summarizes recall records in which the official source identifies Fresenius Medical Care Renal Therapies Group, LLC as the recalling firm or manufacturer. The number of records should not be treated as a safety rating or a measure of company size, because recall counts depend on product mix, reporting periods and how government records are organized.

Compare the dates, categories and classifications together rather than relying on the total alone. Open an individual record for the recalled product, reason, distribution details and current instructions available from the source.