Encore Medical DJO surgical EMPOWR Porous Knee System Empowr 3D Femur-Porous Coated SIZE 8 recall — FDA Z-0508-2024
Due to incorrect product/device within packaging.
Overview of recall records associated with Encore Medical, Lp, including dates, categories and classifications where available.
Classification mix: Class II: 64.
Due to incorrect product/device within packaging.
Knee punch handle tip is breaking during surgery due to the lack of heat treatment, which may lead to a disruption in surgery.
Two different humeral socket Insert devices were swapped during packaging, which could result in surgeons not having the correct device for surgery, which could cause a delay in surgery.
Two different humeral socket Insert devices were swapped during packaging, which could result in surgeons not having the correct device for surgery, which could cause a delay in surgery.
Reverse shoulder prosthesis package that should contain 5mmx30 screws could contain incorrect 3.5mmx30 screws. Use of incorrect screws could result in surgical delay and inadequate fixation.
Knee insert and acetabular system parts were swapped within inner pack prior to the product being sealed within the tray, which could result in surgeons not having the correct product for…
Knee insert and acetabular system parts were swapped within inner pack prior to the product being sealed within the tray, which could result in surgeons not having the correct product for…
Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged…
Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged…
Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged…
Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged…
Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged…
Patella drills are marked with incorrect diameter size of 5.6mm, when they should be 5.4mm. Porous patella overdrills are marked with incorrect diameter size of 5.4 mm, when they should be…
Tibial insert implant large boxes were labeled small and the small boxes were labeled large. If the incorrect labeling is identified this could lead to surgical delay and if the surgeon…
Hip liner implant boxes may contain knee inserts and knee insert boxes may contain hip liner implants, which may lead to implants not being available for surgeries, which may cause delays…
Hip liner implant boxes may contain knee inserts and knee insert boxes may contain hip liner implants, which may lead to implants not being available for surgeries, which may cause delays…
A right, size 8, knee tibial insert package may include a mispackaged, left ,size 4, knee tibial insert. If implanted, a risk may include leaving debris in the joint space.
Complaints have been received concerning the humeral stem failing to mate with the broach handles for a shoulder device. This could result in delay of procedure or revision surgery.
Packaging mixup; There is a potential that packaging identified as 12mm knee prosthesis contains 16mm knee prosthesis and 16mm knee prosthesis packaging contains 12 mm knee prosthesis.
Packaging mixup; There is a potential that packaging identified as 12mm knee prosthesis contains 16mm knee prosthesis and 16mm knee prosthesis packaging contains 12 mm knee prosthesis.
There have been issues experienced with the peg preparation steps, specifically drilling of the Femoral posterior peg. Upon drilling, the shaft portion of the drill may bind within the…
There have been issues experienced with the peg preparation steps, specifically drilling of the Femoral posterior peg. Upon drilling, the shaft portion of the drill may bind within the…
Product mislabeled. Use may result in a delay in surgery, stopping of surgery, and revision surgery.
Due to incomplete machining a small peg remnant on the top of the dome that protrudes above the surrounding porous coating causing the cup not to seat properly.
This page summarizes recall records in which the official source identifies Encore Medical, Lp as the recalling firm or manufacturer. The number of records should not be treated as a safety rating or a measure of company size, because recall counts depend on product mix, reporting periods and how government records are organized.