Dynex Agility Sample Tips recall — FDA Z-2345-2025
The internal label affixed to the sample tip rack is incorrectly labeled as reagent tip racks. When scanned by the Agility instrument, the tip rack is misidentified as a different tip type,…
Overview of recall records associated with Dynex Technologies, Inc., including dates, categories and classifications where available.
Classification mix: Class II: 2.
The internal label affixed to the sample tip rack is incorrectly labeled as reagent tip racks. When scanned by the Agility instrument, the tip rack is misidentified as a different tip type,…
Control samples aspirated from wrong SmartKit on the Agility. Agility software was updated to v1.4.7 to resolve the issue. This leads to a risk that a control from another assay's SmartKit…
This page summarizes recall records in which the official source identifies Dynex Technologies, Inc. as the recalling firm or manufacturer. The number of records should not be treated as a safety rating or a measure of company size, because recall counts depend on product mix, reporting periods and how government records are organized.