Abiomed Automated Impella Controller (AIC) with the below product descriptions and recall — FDA Z-3148-2026
The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
Overview of recall records associated with Abiomed, Inc., including dates, categories and classifications where available.
Classification mix: Class I: 44, Class II: 33.
The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually…
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually…
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually…
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually…
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually…
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually…
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually…
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually…
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually…
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually…
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually…
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually…
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually…
Potential for thrombus formation during prolonged use of the introducer.
Potential for thrombus formation during prolonged use of the introducer.
Potential for thrombus formation during prolonged use of the introducer.
Potential for thrombus formation during prolonged use of the introducer.
Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.
Potential software error in the Automated Impella Controller ("AIC" ) when used in conjunction with left ventricular Impella devices.
Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC…
A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through…
This page summarizes recall records in which the official source identifies Abiomed, Inc. as the recalling firm or manufacturer. The number of records should not be treated as a safety rating or a measure of company size, because recall counts depend on product mix, reporting periods and how government records are organized.
Compare the dates, categories and classifications together rather than relying on the total alone. Open an individual record for the recalled product, reason, distribution details and current instructions available from the source.